Thon ring patterns with FEI Falcon and Gatan US4000
Forums
We are testing our new direct electron detectors (FEI Falcon and Gatan K2 Summit). Rachel Ruskin in the lab started comparing Thon ring patterns (see attachments) calculated from images of amorphous carbon film using the Falcon detector and Gatan US4000 CCD (scintillator-based):
Microscope: FEI TF20 Voltage: 200 kV Magnification: 29,000 Dose rate: Falcon: 30 electrons/Å2/second US4000: 25 electrons/Å2/second Exposures: Falcon: 0.4 sec, 1.0 sec, 5.0 sec, 7.5 sec US4000: 0.4 sec, 1.0 sec, 5.0 sec, 10.0 sec, 20.0 sec, 40 sec Dose/pixel: Falcon: 12, 30, 150, 225 electrons US4000: 10, 25, 125, 250, 500, 1000 electrons
Although the dose values are not exactly the same in these two exposure series it is clear that the Falcon shows Thon rings that extend all the way to Nyquist (see 7.5 second exposure) while the stop about 0.8 Nyquist in a similar image recorded using the US4000 (see 10.0 second exposure).
We will perform more tests, including the K2 Summit and film and will also perform quantitative assessments by measuring DQEs.
Thon rings from K2 summit at 200kV
Dear Niko,
Nice to see the comparison here. Do you have the FFTs to show Thon rings for K2 summit under the similar conditions (200kV, 29kx and different exposures) you posted here?
Thanks in advance!
Dong-Hua
I a sorry but I do not have
In reply to Thon rings from K2 summit at 200kV by donghua
I a sorry but I do not have them. This is rather old data and I am not even sure I could find them in my backups anymore.